Manufacturer: B.I.A IMPEX Pakistan.
DATE LAST AMENDED: 18 Oct. 2004
DESCRIPTION: ElectroSurgery Bipolar &
Monopolar Cables 3m & 5m
Materials used in Manufacture:
|
|
Connectors: |
Santoprene Thermoplastic Elastomer (
Medical Grade)
|
|
Cable: |
Silicone insulated 5.2mm x 2.5mm with
384 strands of 0.05 tinned copper wire
for
Bipolar Cable.
|
|
Contacts: |
Nickel-plated brass or Stainless Steel
|
|
Electrical performance & suitability: |
The product has been designed in
accordance with relevant standards
and is suitable for use with all
suitable Electro-Surgery generators and
connective instruments. All products
are latex free in construction.
|
TECHNICAL SPECIFICATION:
Product Packaging & labeling: The product is
packaged in a polythene bag to maintain
cleanliness and
must be Autoclaved prior to use ( Except
Footswitches & Plate cables). The label has the
Stock code,
Description, batch no. and supplier details for
identification.
Contacts:
Nickel-plated brass or Stainless Steel
Electrical performance & suitability: The
product has been designed in accordance with
relevant standards
and is suitable for use with all suitable
Electro-Surgery generators and connective
instruments. All products
are latex free in construction.
GENERAL INSTRUCTIONS FOR USE: -
Follow recommended guidelines for power set up &
connection to generator and make sure instrument
is
pushed fully home into its socket prior to
switching on. Always make sure adhesive plate
connections are
sound and secure to prevent any chance of burns
in this area.
Recommended Disposal Method:
The product has no
user serviceable parts and should be incinerated
with other medical waste. Every care is taken to
ensure negative toxicity of all materials used
and
prolonged contact with skin will not cause any
adverse reactions.
CLEANING/ STERILIZATION METHOD:
Cleaning General: To remove blood & general
grime the product should be cleaned in a mild
detergent
prior to sterilization. DO NOT use solvents of
any as this might affect plastic parts. The
product may be
cleaned using an Instrument cleaner using a 5
cycle process ( Pre-rinse, wash, disinfect, R O
rinse & dry ) if
this is accepted practice.
Autoclaving:
Wrapped Sterilization at 115-138°c
for 3-4 minutes with post drying and make sure
heavy
objects are not placed on the product during and
whilst cooling down. Some Autoclaves feature a
Prion
cycle; this may be used if required
Ethylene Oxide: This can be used if all traces
are removed by post drying prior to use.
Radiation:
The product can be exposed to Cobalt 60
Gamma radiation at up to 4 megarads although a
level
of 2-3 is quite sufficient for thorough
sterilization.
PRODUCT LIFE & INSPECTION CRITERIA: -
The product can be used for a considerable
amount of time if it is routinely inspected: -
A) Check plated metal contacts; these should be
bright without corrosion.
B) Plastic connectors should be free from
imperfection & distortion.
C) Cable insulation must be checked
periodically to make sure the product is safe
for use, pay special
attention to the exit point on the connectors as
excessive bending can cause failure in this
area.
D) All electrical cables and instruments must
be regularly checked for electrical continuity
by the
Medical Physics Department to prevent possible
failure whilst being used in surgery. If this is
not
possible continuity testers are available from
your supplier for on site testing. |